Omega CL526 Binding Machine User Manual


 
9
1.2.1 Table of ranges and accuracy
Sensor or Total range High accuracy Resolution Limit of error
parameter range
Tc type J -210 to +1200°C -190 to +1200°C 0.1°C ± (0.02% of rdg +0.1°C)
-346 to +2192°F -310 to +2192°F 0.1°F ± (0.02% of rdg +0.18°F)
Tc type K -270 to +1370°C -150 to 1300°C 0.1°C ± (0.02% of rdg +0.1°C)
-454 to +2498°F -220 to 2372°F 0.1°F ±(0.02% of rdg +0.18°F)
Tc type T -270 to +400°C -150 to 400°C 0.1°C ± (0.02% of rdg +0.1°C)
-454 to +752°F -202 to 752°F 0.1°F ±0.02% of rdg +0.18°F)
Tc type R 0 to +1760°C 500 to 1700°C 0.1°C ± (0.02% of rdg +0.3°C)
32 to +3200°F 932 to 3092°F 0.1°F ± (0.02% of rdg +0.154°F)
Tc type S 0 to +1760°C 600 to 1760°C 0.1°C ± (0.02% of rdg +0.3°C)
32 to +3200°F 1112 to 3200°F 0.1°F ± (0.02% of rdg +0.154°F)
Tc type B 200 to +1820°C 1000 to 1820°C 0.1°C ±(0.02% of rdg +0.4°C)
392 to +3308°F 1832 to 3308°F 0.1°F ± (0.02% of rdg +0.172°F)
Tc type C 0 to +2300°C 1150 to 2300°C 0.1°C ± (0.02% of rdg +0.4°C)
32 to +4172°F 2102 to 4172°F 0.1°F ± (0.02% of rdg +0.172°F)
Tc type G 0 to +2300°C 300 to 2000°C 0.1°C ± (0.02% of rdg +0.4°C)
32 to +4172°F 572 to 3632°F 1°F ± (0.02% of rdg +0.172°F)
Tc type D 0 to +2300°C 300 to 2000°C 0.1°C ± (0.02% of rdg +0.4°C)
32 to +4172°F 572 to 3632°F 0.1°F ± (0.02% of rdg +0..72°F)
Tc type U -200 to +400°C -150 to +400°C 0.1°C ±(0.02% of rdg +0.1°C)
-328 to +752°F -238 to +752°F 0.1°F ± (0.02% of rdg +0.18°F)
Tc type L -200 to +760°C -200 to 760°C 0.1°C ± (0.02% of rdg +0.1°C)
-328 to +1400°F -328 to 1400°F 0.1°F ± (0.02% of rdg +0.18°F)
Tc type N 0 to 1300°C 0 to 1300°C 0.1°C ±(0.02% of rdg +0.1°C)
+32 to 2372°C 32 to 2372 0.1°F ± (0.02% of rdg +0.18°F)
Tc type E -270 to +1000°C -200 to +1000°C 0.1°C ±(0.02% of rdg +0.1°C)
-454 to +1832°F -328 to +1832°F 0.1°F ± (0.02% of rdg +0.18°F)
Tc type F 0 to 1400°C 0 to 1400°C 0.1°C ±(0.02% of rdg +0.1°C)
32 to 2552°F 32 to 2552°F 0.1°F ± (0.02% of rdg +0.18°F)
Pt 100 (IEC) -200 to +850°C -200 to +850°C 0.1°C ±(0.02% of rdg +0.1°C)
-328 to +1562°F -328 to +1562°F 0.1°F ± (0.02% of rdg +0.18°F)
Pt100 (JIS) -200 to +600°C -200 to +600°C 0.1°C ±(0.02% of rdg +0.1°C)
-328 to +1562°F -328 to +1562°F 0.1°F ± (0.02% of rdg +0.18°F)
Pt100 (US) -200 to +850°C -200 to +850°C 0.1°C ±(0.02% of rdg +0.1°C)
-328 to +1562°F -328 to +1562°F 0.1°F ± (0.02% of rdg +0.18°F)
Ni100 -60 to +180°C -60 to +180°C 0.1°C ±(0.02% of rdg +0.1°C)
-76 to +356°F 0.1°F ± (0.02% of rdg +0.18°F)
Ni120 0 to +150°C 0 to +150°C 0.1°C ±(0.02% of rdg +0.1°C)
32 to +302°F 32 to +302°F 0.1°F ± (0.02% of rdg +0.18°F)
mV -18.000 to +22.000 -10.000 to +22.000 1 µV ± (0.01% of rdg +3µV)
mV -10.00 to +100.00 -10.000 to +21.000 10 µV ± (0.01% of rdg +3µV)
+21.00 to +53.00 10 µV ± (0.01% of rdg +3µV)
+53.00 to +100.00 10 µV ± (0.01% of rdg +6µV)
mV 0.0 to +1000.0 0.0 to +1000.0 100 µV ± (0.01% of rdg +40µV)
V 0.000 to +10.000 0.000 to +10.000 1 mV ± (0.02% of rdg +0.4mV)
mA 0.000 to +21.000 0.000 to +21.000 1 µA ± (0.02% of rdg +0.5 µA)
(IN) 0.0 to 400.0
0.0 to 400.0
10 m
± (0.02% of rdg +38m
)
(OUT) 0.0 to 400.0
0.0 to 400.0
10 m
± (0.03% of rdg +78m
)
Note:
Accuracy shown are based on tests at 23°C ±5°C for 360 days
All Input ranges: additional error ± 1 digit
Traceability chart to WECC or SIT available on request