( 43 / 72 )
4.3 Self-checks
(1) Self-check Procedure
To run the emulator self-checks, do so as explained here below. While self-checks are in progress,
LEDs will change as shown in Figure 4.4.
(1) Set the switches in the emulation pod same as the factory setting (see Figure 4.3).
(2) When the target system is connected, disconnect it.
(3) Within 2 seconds of activating power to the emulator, press the RESET switch on the emulator
front panel to switch the emulator to the maintenance mode.
(4) Check the "SAFE" LED starts blinking and then press the RESET switch again.
(5) The self-checks will start. If the normal result is displayed in about 2 minutes, self-checks
terminated normally.
Figure 4.3 Switch settings when running self-checks (factory-settings)
(2) If an Error is Detected in Self-checks
If self-checks do not result normally (ERROR 1 and ERROR 2 in Figure 4.4), check the following.
• Check the connection of the emulation pod and the PC4701.
• Download the proper firmware.
• Check if the switches in this product are set same as the factory-settings (Figure 4.3).
CAUTION
Note on Self-checks:
• If self-checks do not result normally (excluding target system errors), the emulation
pod may be damaged. Then contact your local distributor.