Sigma BM-286A Computer Monitor User Manual


 
3. OPERATING INSTRUCTIONS page 14
Model BM-286A
3.4.1. Set-Up Menu:
Set-up is used to refer to the parameters that are common to all waste types.
The menu will look something like this:
SET-UP MENU
Bkg Time 120
Bkg Method Rolling
Low Alarm 0
High Alarm 10000
Var Time 60
Var Trials 3
Units of Weight Lbs
Min Weight lbs 10
Reports All
BM I.D.: ABCDEFGH
User Stream
0
OK
The available menu parameters are defined as follows:
Background Time:
Background time, in seconds. Enter a value between 20 and 10,000.
Background Method:
Two methods of background count are incorporated in the BM-286A. They are
selectable from the Set-Up Menu. The Rolling background method is the
standard for TSA products. For a count time of N seconds, the count is updated
every N/10 seconds, and the sum of the last 10 count periods are averaged to
determine the current background. The Standard background method simply
counts for N seconds.