Soehnle 9220 Scanner User Manual


 
15
ENGLISH
Calibration
To enter the "calibration" mode, press any key at the same
time as switching on the instrument.
Press the key for scale A or the key for scale B.
A number between 5,000 and 50,000 is displayed.
If this is not the case, the scale must be re-calibrated.
If the display is stable, press the key to set the scale to
zero.
Then place the calibration weight (3/6 kg = 500 g,
30 kg = 1000 g) on the platform.
(display 200000 ... 800000 at max. weight).
Enter the value of the calibration weight (3/6 kg = 500 ,
30 kg = 1) and press the key to calibrate the scale.
Switch off the scale, remove the load and switch the scale
on again.
The scale is ready for use.
Error messages
Error
Max. weighing range
exceeded.
Weighing platform not cor-
rectly mounted. Scale outsi-
de set-to-zero range.
Set to zero not possible.
Battery almost empty.
Remedy
Remove weighing product
from scale!
Mount platform correctly!
Make sure that the scale is
empty!
Connect the power plug!