Soehnle 9222 Postal Equipment User Manual


 
11
Programming the Tare Weight
as Standard
Place the tare weight on the scales before
switching them on.
Switch on the scales.
Continue as described in "Saving PLU
(Product Look Up)".
Error Messages
Max. weighing range exceeded
> Take the items off the scales
Piep Zero setting not possible
> Deviation from zero point too high
User Calibration
• Press and switch on the scales with the
On
key.
Select using
C
"CAL 4"
and confirm using
.
Selecting the maximum weight
Select the maximum weight in kg
(2nd row) using
C
and confirm
with
.
Zero point calibration
< A/D reading
< reading 0 (grams)
< "Pnt 0"
Make sure that the scales are empty.
Wait until the A/D reading (1st row) is stable,
then press
to continue.
Setting the calibration weight
< A/D reading
< calibration weight 0 (grams)
< "Pnt. CAL"
Place the calibration weight (min. 1/3 of the
weighing range) on the platform. Enter the
weight in grams via the keypad. Wait until
the A/D reading is stable, then press
.
The calibration was successful.
Error Message "Err. Ad"
> Calibration weight too low
> Deviation from calibration weight
too high
Use a calibration weight with at least 1/3 of the
weighing range. You should calibrate in the range
in which you need to achieve the highest degree
of accuracy.
The power-up initialisation starts when the
scales are discharged.