4000 Configuration
3.4.3 Configuration Menu
Configuration Menu Description
Level Units Level units used for display and Modbus data. FT-IN-16TH, M,
or FT are valid. Use F1/F2 to select.
Temp Units Temperature units used for display and Modbus data. F, C, or K
are valid. Use F1/F2 to select.
Level Sensor Direction of gauge. Forward/Reverse
Calib Level2 Command used to calibrate to a new level. Pressing F4 causes
ATT to ask for the current level to be entered.
Batt Off2 Command used to turn the battery off. Pressing F4 causes the
battery to be turned off.
Damping Damping value used for level and temperature readings. 0-16
is valid.
Temp Sensor Type of temperature sensor attached. None, Copper RTD, or
Platinum RTD are valid. Use F1/F2 to select.
Manual Temperature If there is no temperature sensor, this is the value displayed
and returned for temperature.
Noise Reject Sets up the noise rejection for the A/D. 50 and 60 Hz are valid.
Use F1/F2 to select.
Critical High Level Critical high level alarm setpoint.
Advisory High Level Advisory high level alarm setpoint.
Advisory Low Level Advisory Low LevelAdvisory low level alarm setpoint.
Critical Low Level Critical low level alarm setpoint.
Level Deadband Deadband used for level alarms.
Critical High Temp Critical high temperature alarm setpoint.
Advisory High Temp Advisory high temperature alarm setpoint.
Advisory Low Temp Advisory low temperature alarm setpoint.
Critical Low Temp Critical low temperature alarm setpoint.
Temp Deadband Deadband used for temperature alarms.