A SERVICE OF

logo

268
Programming Commands
ML Mixer Level
NOTE If an external amplifier gain value is set, the mixer level is determined using the
following equation:
Mixer Level = Ref. Level - Attenuation + Ext. Amplifier Gain
The external amplifier gain is not preset by doing an IP command in case the
analyzer is measuring a large signal. This is to protect the analyzer from damage
from a large signal. For a helpful suggestion, see Chapter 3 , “Hints and Tips,” on
page 71.
NOTE For 8566A/B, 8568A/B, the functions of the ML command are identical to the KS,
command (page 179).