8-6. Correcting the Internal Mass Value
The balance can correct the internal mass value within the range shown below. This
function corrects the internal mass value to conform to an external weight. The corrected
mass value is maintained in non-volatile memory even if the AC adapter is removed. The
internal mass value is corrected as follows:
Model Target Range
MC-10K 2 kg ±0.500 g
MC-30K 10 kg ±5.00 g
Step 1 Calibrate the balance using the internal mass. (one-touch calibration).
Place the external weight and find out the correction value.
Example for correcting the weight value by +0.30 g in 10 kg, when
using MC-30K. If correcting the weight value by +0.30 g in 5 kg,
the weight changed into 10 kg, the correction value is +0.60 g.
Step 2 Press the ON:OFF key to turn off the display.
Step 3 While pressing and holding the PRINT key and
the SAMPLE key, press the ON:OFF key. The
balance displays p5 .
Step 4 Press the PRINT key. Then the balance displays the function
switches. Set the function table switch and internal mass
correction switch to "1" as shown above using the following keys.
SAMPLE key
...
To select the switch to change the value.
The selected digit blinks.
RE-ZERO key
..
To change the parameter of the switch selected.
Step 5 Press the PRINT key to store the new setting.
The balance returns to the weighing mode.
Step 6 Press and hold the
SAMPLE key to enter the function
table and release the key when ba5fnc is displayed.
Step 7 Press the SAMPLE key several times until C5 in is
displayed, then release the key.
Step 8 Press the PRINT key to enter the procedure for
correcting the internal mass value.
Step 9 Correct the internal mass value using the following keys.
RE-ZERO key
..
To select the value. (-500 digits appear
after +500 digits)
PRINT key
.......
To store the new value and display the
next menu item of the function table.
Correction the internal
mass by +0.30 g at 10 kg.
Calibrate with the
corrected internal mass.
The same mass
1
0 kg
1
0 kg
1
0000
.
00
g
1
0000
.
30
g
Corrected external weigh
t
Function table switch
Correction switch of internal mass
0. 30
g
C5 in
ba5fnc
0. 0
g
a-1
xxx
1
a-01111
p5
With these keys Press the key
held down
Press and hold