KF format 5if type 2
This is the Karl-Fischer moisture meter format and is used when the peripheral equipment
can only communicate using this format.
This format consists of fourteen characters excluding the terminator.
This format has no header characters.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data
is not zero or overloaded.
This format outputs the unit only for a stable value.
MT format 5if type 3
A header of two characters indicates the balance condition.
The polarity sign is used only for negative data.
The weighing data uses spaces in place of the leading zeros.
The character length of this format changes dependent upon the unit
NU (numerical) format 5if type 4
This format outputs only numerical data.
This format consists of nine or ten characters excluding the terminator.
With the MC-10K, when the numerical data excluding the decimal point exceeds eight
characters, the data format will be ten characters long.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is used.
(Data example exceeding 8 characters)
CSV format 5if type 5
This format separates the data of A&D standard format and the unit by a comma ( , ).
This format outputs the unit even when the data is overloaded.
When the ID number, data number, time and date are added at "Data output (dout)" of the
function table, outputs ID number, data number, date, time and weighing data in this order
and separates each item by a comma and treats all the items as one group of data.
g
+10000.000 g
C
R
L
F
Data
Unit
Terminator
Stable value
Unstable value
S 10000.000 g
C
R
L
F
S
SD
SI
Stable header
Unstable header
Overload header
Header Unit Terminator
Data
C
R
L
F
ST,+10000.000, g
C
R
L
F
OL,+9999999E+19,
g
LAB-123, No,012, 2009/12/31, 12:34:56, ST,+10000.000, g<CR><LF>
ID number Data number Date Time Weighing data
+0000.127
C
R
L
F
Data
Terminator
+10000.000
C
R
L
F
Data
Term ina tor