A&D MC-10K Postal Equipment User Manual


 
49
14. Extended Function
The MC series balance has several extended functions equipped for special applications or to
troubleshoot when using the standard functions.
When the "Filter ( f1l)” is set to “ 0 “, extra items are available as shown below.
Class Item and Parameter Description
0
1
2
3
4
f1-b
Averaging range for
the first moving
average
5
Small
Large
When the fluctuation of a weight value is within
the averaging range, the averaging operation
starts to stabilize the displayed value.
When the fluctuation of a weight value is
small, for example, when weighing or filling a
small amount of sample, the averaging is
always performed and the response rate
maybe slow. Under such a condition, change
the parameter.
Refer to “Averaging range
(
f1-b
)
and
averaging time
(
f1-t
)”
0
No averaging
1
0.5 second
2
1.0 second
3
1.5 seconds
4
2.0 seconds
5
2.5 seconds
f1-t
Averaging time for
the first moving
average
6
3.2 seconds
When the fluctuation of a weight value is within
the averaging range, the averaging operation
starts. Once the duration of averaging operations
reach the averaging time, moving averaging
starts. This parameter sets the time to start
moving averaging.
Refer to “Averaging range (
f1-b
) and
averaging time (
f1-t
)”
0
1
2
3
4
f2-b
Averaging range for
the second moving
average
5
Small
Large
Refer to “Filter depending on differences in
the amount to deliver powdery and liquid
material”.
0
No averaging
1
0.5 second
2
1.0 second
3
1.5 seconds
4
2.0 seconds
5
2.5 seconds
erfnc
Extended
function
f2-t
Averaging time for
the second moving
average
6
3.2 seconds
:
Factory settings.