External source mode
1
(CW time sweep only)
Frequency range 300 kHz to 6 GHz
R-channel input requirements (typical)
Power level 0 to -25 dBm
Spectral purity
Maximum <-30 dBc
spurious input
Residual FM <20 kHz
Typical settling time 500 ms (automatic)
50 ms (manual)
Frequency readout accuracy 0.1% (automatic)
Input frequency accuracy requirement
2
Manual -0.5 to 5 MHz
Accuracy (see magnitude and phase
characteristics)
Magnitude Characteristics
Dynamic accuracy (10 Hz IF BW)
Display resolution 0.001 dB/division
Marker resolution
3
0.001 dB
Trace noise
(+5 dBm at test-port, ratio measurement, 3 kHz BW)
300 kHz to 3 GHz <0.006 dB rms
3 GHz to 6 GHz <0.010 dB rms
Reference level
Range ±500 dB
Resolution 0.001 dB
Stability
2
30 kHz to 3 GHz 0.02 dB/° C
3 GHz to 6 GHz 0.04 dB/° C
Phase Characteristics
Dynamic accuracy (10 Hz IF BW)
Range ±180°
Display resolution 0.01°/division
Marker resolution
3
0.01°
Trace noise
(+5 dBm at test-port, ratio measurement, 3 kHz BW)
300 kHz to 3 GHz <0.038° rms
3 GHz to 6 GHz <0.070° rms
Reference level
Range -180° to +180°
Resolution 0.01°
Stability
30 kHz to 3 GHz 0.05°/° C
3 GHz to 6 GHz 0.20°/° C
Polar characteristics
Range 10 x 10
–12
to 1000 units
full scale
Reference ±500 units
1. See the Agilent 8753ES descriptions and options for a functional description.
Measurement accuracy is dependent on the stability of the input signal.
2. Typical performance.
3. Marker resolution for magnitude; phase and delay is dependent upon measured
value. Resolution is limited to five digits.