Agilent Technologies 8753ET Network Router User Manual


 
15
External source mode
1
(CW time sweep only)
Frequency range 300 kHz to 6 GHz
R-channel input requirements
2
Power level 0 to -25 dBm
Spectral purity
Maximum spurious input <-30 dBc
Residual FM <20 kHz
Typical settling time 500 ms (automatic)
50 ms (manual)
Frequency readout accuracy
2
0.1% (automatic)
Input frequency accuracy requirement
2
Manual -0.5 to 5 MHz
Accuracy (See magnitude and phase
characteristics)
Magnitude characteristics
Dynamic accuracy (10 Hz IF BW)
Display resolution 0.001 dB/division
Marker resolution
3
0.001 dB
Trace noise
(0 dBm at transmission port or +5 dBm at reflection port,
ratio measurement, 3 kHz BW)
300 kHz to 3 GHz <0.006 dB rms
3 GHz to 6 GHz <0.010 dB rms
Reference level
Range ±500 dB
Resolution 0.001 dB
Stability
2
300 kHz to 3 GHz 0.02 dB/1° C
3 GHz to 6 GHz 0.04 dB/1° C
Phase Characteristics
Dynamic accuracy (10 Hz IF BW)
Range ±180°
Display resolution 0.01°/division
Marker resolution
3
0.01°
Trace noise
(0 dBm at transmission port or +5 dBm at reflection
port, ratio measurement, 3 kHz BW)
300 kHz to 3 GHz <0.038° rms
3 GHz to 6 GHz <0.070° rms
Reference level
Range -180° to +180°
Resolution 0.01°
Stability
30 kHz to 3 GHz 0.05°/° C
3 GHz to 6 GHz 0.20°/° C
Polar characteristics
Range 10 x 10
–12
to 1000 units full
scale
Reference ±500 units
1. See the Agilent 8753ET/ES descriptions and options for a functional description.
Measurement accuracy is dependent on the stability of the input signal.
2. Typical performance.
3. Marker resolution for magnitude, phase and delay is dependent upon measured
value. Resolution is limited to five digits.