Agilent 8753ES (75-ohm systems)
Type-F test ports
Configuration
Network analyzer 8753E Option 075
Calibration kit 85039B
Test-port cables 11857B
System dynamic range
30 kHz to 50 kHz 68 dB
1
50 kHz to 300 kHz 90 dB
1
300 kHz to 16 MHz 96 dB
16 MHz to 3 GHz 106 dB
Measurement Port Characteristics
Data is shown for a Type-F female reflection port and
a Type-F male transmission port.
Frequency Range
Corrected
2
30 kHz–300 kHz 300 kHz–1.3 GHz 1.3 GHz–3 GHz
Directivity 45 dB 45 dB 40 dB
Source match 40 dB 40 dB 30 dB
Load match 45 dB 45 dB 40 dB
Reflection ±(0.060 dB ±(0.060 dB ±(0.024 dB
tracking +0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C)
Transmission ±(0.033 dB ±(0.019 dB ±(0.057 dB
tracking +0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C)