Agilent 8753ES (50-ohm systems)
7-mm test ports
Configuration
Network analyzer 8753ES
Standard and Option 006
Calibration kit 85031B
Test-port cables 11857D
System dynamic range
30 kHz to 50 kHz 70 dB
1
50 kHz to 300 kHz 90 dB
1
300 kHz to 16 MHz 100 dB
16 MHz to 3 GHz 110 dB
Option 014 108 dB
3 GHz to 6 GHz 105 dB
Option 014 103 dB
Measurement Port Characteristics
Frequency Range
Corrected 30 kHz– 300 kHz– 1.3 GHz– 3 GHz–
300 kHz
2
1.3 GHz 3 GHz 6 GHz
Directivity 55 dB 55 dB 51 dB 46 dB
Source match 55 dB 51 dB 49 dB 43 dB
Load match 55 dB 55 dB 51 dB 46 dB
Reflection ±(0.001 dB ±(0.001 dB ±(0.005 dB ±(0.020 dB
tracking +0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)
Transmission ±(0.008 dB ±(0.006 dB ±(0.009 dB ±(0.021 dB
tracking +0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)
Uncorrected
Directivity 20 dB* 35 dB 30 dB 25 dB
Source match 18 dB** 16 dB 16 dB 14 dB
Load match 18 dB** 18 dB 16 dB 14 dB
Reflection ±2.5 dB ±1 dB ±1 dB ±1.5 dB
tracking
Transmission ±2.5 dB ±1 dB ±1 dB ±1.5 dB
tracking
Crosstalk 90 dB*** 100 dB 100 dB 90 dB
*15 dB, 30 kHz to 50 kHz **10 dB, 30 kHz to 50 kHz ***60 dB, 30 kHz to 50 kHz
System Performance Summaries
1. Typical below 300 kHz.
2. Typical performance.
Reflection uncertainty
Magnitude Phase
Magnitude Phase
Transmission uncertainty