Agilent Technologies G3180B Computer Accessories User Manual


 
25
Agilent G3180B Splitter Kit
Installation and Operation Guide
Agilent Technologies
3
Splitter Configurations
Typical Configurations 26
Splitting to an MSD 28
Custom Configurations 29
Restrictor id and length 32
Maximum and minimum flows 33
Column outlet pressure 34
Inlet pressure 34
Restrictor and Column Installation 35
Install the column 35
Connect the splitter 35
Disconnect tubing from the splitter 37
The combination of restrictor diameters and lengths determines how the
column effluent is divided (the split ratio) between the two detectors. There
are two approaches to setting up a splitter method.
Use a typical configuration. A set of eight configurations is discussed
beginning on the next page. They apply to a variety of detector
combinations and split ratios. All of the flows have been calculated.
Create a custom configuration. If the typical configurations do not meet
your needs, you can create one that does. The CD shipped with the splitter
kit provides tools for the necessary calculations.
We suggest examining the typical configurations first, since they cover a wide
variety of splitter applications and require no calculations.