Agilent Technologies N5700 Computer Accessories User Manual


 
Calibration
112 Series N5700 User’s Guide
*OPC?
Step 4. Measure the output voltage with the Agilent 3458A and enter the
data.
CAL:DATA <data>
Step 5. Select the second voltage calibration point.
CAL:LEV P2
*OPC?
Step 6. Measure the output voltage with the Agilent 3458A and enter the
data.
CAL:DATA <data>
Current Programming and Measurement Calibration
Step 1. Connect a precision shunt resistor to an output. Connect the Agilent
3458A to measure the voltage across the shunt. The shunt resistor
should be suitable for measuring at least 120% of the power supply’s
rated full-scale current.
Step 2. Select the full-scale current programming range. The value to
program a range must be the maximum current of the range.
CAL:CURR <current>
Step 3. Select the first current calibration point.
CAL:LEV P1
*OPC?
Step 4. Calculate the shunt current (I=V/R) and enter the data.
CAL:DATA <data>
Step 5. Select the second current calibration point.
CAL:LEV P2
*OPC?
Step 6. Calculate the shunt current (I=V/R) and enter the data.
CAL:DATA <data>
Exit Calibration mode
CAL:STAT OFF