Barco 10MP Computer Monitor User Manual


 
34 Coronis Fusion 10MP
Reading rooms
ABOUT READING ROOMS
The American Association of Physicists in Medicine (AAPM) composed a
list of pre-defined reading rooms. Each of these reading rooms are
defined by following parameters:
the maximum light allowed in this type of room
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset
ambient light correction value to take into account to recalculate the
DICOM display function when Ambient Light Compensation (ALC) is
enabled. Please refer to “Ambient Light Compensation (ALC)” on
page 33 to enable ALC.
The available reading rooms for your C
ORONIS FUSION 10MP display are:
Reading rooms can only be selected when the DICOM display
function is selected. Therefore, please refer to “Display
functions” on page 32 to correctly set the display function.
CR/DR/MAMMO Corresponds to light conditions in
diagnostic reading rooms for computed
radiology, digital radiology or
mammography. This setting has the lowest
maximum ambient light.
CT/MR/NM Corresponds to light conditions in
diagnostic reading rooms for computed
tomography, magnetic resonance or
nuclear medicine scans.
Staff Office Corresponds to light conditions in office
rooms.
Clinical Viewing
Room
Corresponds to light conditions in
diagnostic reading rooms for clinical
viewing.