To perform the calibrations:
1 Turn the control wheel to display the Calibration menu.
2 Click the control wheel to enter the Calibration menu.
3 Turn the control wheel to select the calibration you wish to
perform.
4 For the automatic calibrations: Click the control wheel to perform
the calibration. During the calibration, the progress bar in the on-
screen display is filled out.
After successful calibration, the on-screen display shows the word
"Successfully".
After unsuccessful calibration, the on-screen display shows the
word “Failed”. Please refer to chapter 5 : Troubleshooting for
more information.
5 For manual calibrations:
a) Click the control wheel to enter the menu.
b) Turn the control wheel to change the contrast or brightness
value.
c) Click the control wheel to confirm the change of the value.
6 After calibration, click and hold the control wheel for a few
seconds to exit the Calibration menu.
The different calibration functions:
Total Autoset ............This function performs the automatic detection
of the number of pixels (horizontal and vertical)
and all other calibration functions (described
below) one after the other. This function takes a
few minutes.
Phase & level ...........Automatic calibration of the video sampling
phase and the video levels (black and white)
Auto level..................Automatic calibration of the video levels (black
and white)
Auto phase ...............Automatic calibration of the video sampling
phase
Manual phase...........Manual adjustment of the video sampling phase
Manual black ............Manual adjustment of the video black level
(brightness cal setting)
Manual white ............Manual adjustment of the video white level
(contrast cal setting)