B&K 1787B Power Supply User Manual


 
13
Voltage
<0.05% + 10mV
Programming
Accuracy, 12months
,
(
@ 25 °C ± 5 °C)
±
(% of output+offset)
Current
<0.2% +10mA
Voltage
<0.05%+15mV(<20V),
<0.05%+120mV(=20V)
Readback
Accuracy
12months
,
(25 °C ± 5 °C)
±(% of output+offset)
Current
<0.1% +15mA
Voltage
=1mVrms/3mVp-p
Ripple
(20Hz ~20MHz)
Current 5mArms
Voltage
<0.02%+5mV
Tem
perature
C
oefficient,
(0 °C ~ 40 °C)
±(% of output+offset)
Current
<0.1% +5mA
Voltage
<0.02%+15mV(<20V),
<0.02%+120mV(=20V)
Read
back
Y
emperature,
Coefficient,
±(% of output+offset)
Current
<0.1% +15mA
2.2 Supplemental Characteristics
State Storage Memory
Sixteen (16) user-configurable stored states
Recommended Calibration Interval
1 year
AC Input Ratings (selectable via switch on the rear panel)
Option OP1: 220VAC ± 10%, 47 to 63 Hz
Option OP2: 110 VAC ± 10%, 47 to 63 Hz
Maximum input power
350VA
Cooling
Fan cooled
Operating Temperature
32 to 104 °F (0 to 40 °C) for full rated output
Storage Temperature
-68 to 158 °F (-20 to 70 °C) for storage environment.