Emerson NGA 2000 Computer Accessories User Manual


 
2 - 14
AK
90003752(1) [AK-Commands] 10/98
Stop command
STBY K0
Function associated to the whole system unit
Code
Only using the channel number 0 (K0) will stop running "SYSCAL" procedure. Besides, all
procedures of the other analyzer modules will be stopped.
Response
SCAL 0
Error status
Code
Read command
ASTZ K0
Read of the whole FU
Code
With the command "ASTZ K0" it will be checked, if a system calibration is running or not.
"SCAL" will be sent back for a running system calibration. If no system calibration is
running this string will be missed.