FieldServer IR-148 Computer Drive User Manual


 
FS-8700-102 Honeywell Zellweger IR-148-8_8-Channel Manual Page 11 of 19
FieldServer Technologies 1991 Tarob Court Milpitas, California 95035 USA Web:www.fieldServer.com
Tel: (408) 262-2299 Fax: (408) 262-9042 Toll_Free: 888-509-1970 email: support@fieldServer.com
5.2. Server Side Node Descriptors
Section Title
Nodes
Column Title Function Legal Values
Node_Name Provide name for node
Up to 32 alphanumeric
characters
Node_ID
Commonly used parameter but
irrelevant for this driver as
Protocol is node-less.
A unique Node_ID is, however,
recommended for each
Honeywell Zellweger IR-148
Unit as it may be useful in
exposing Node status
information
1-255
Protocol Specify protocol used InfraTox, Infra-Tox.
Port
Specify at which port the device
is connected to the FieldServer
Ports P1-P8 can be used with
232/485 converter.
P1-P8 (with 232/485
converter), R1-R2
Example
// Server Side Nodes
Nodes
Node_Name, Node_ID, Protocol, Port
DEVR1, 1, InfraTox, R1
5.3. Server Side Map Descriptors
5.3.1. FieldServer Specific Map Descriptor Parameters
Column Title Function Legal Values
Map_Descriptor_Name
Name of this Map
Descriptor
Up to 32 alphanumeric
characters
Data_Array_Name
Name of Data Array where
data is to be stored in the
FieldServer
One of the Data Array
names from “Data Array”
section above
Data_Array_Offset
Starting location in Data
Array
0 to maximum specified in
“Data Array” section above
Function
Function of Server Map
Descriptor
Wrbc