Fluke 45 Printer User Manual


 
Maintenance
Replacement Parts
6
6-7
Table 6-3. Performance Tests for mA Current Functions
Display
Function
Range (Rate) Input Level
Frequency
Min Max
[ (mA)
\ (mA)
30 mA
100 mA
30 mA
100 mA
+30 mA
+100 mA
30 mA
100 mA
1 kHz
1 kHz
29.982
99.93
29.840
99.40
30.018
100.07
30.160
100.60
Table 6-4. Performance Tests for A Current Functions
Display
Function
Range (Rate) Input Level
Frequency
Min Max
[
\
10 A
10 A
+10 A
10 A
1 kHz
9.975
9.890
10.025
10.110
Replacement Parts
Parts that can be replaced by the user are listed in Table 6-5. Other parts must be replaced
by qualified service personnel only, following the procedures in the Service Manual.
Replacement parts and price information can be obtained from Fluke or an authorized
Fluke distributor. To ensure prompt and efficient handling, include the following
information with your order:
Quantity
Fluke Part Number
Part Description
Instrument Model and Serial number.
Table 6-5. Replacement Parts
ITEM DESCRIPTION FLUKE PART NO.
C40*
F1 W
F2 W
F3 W
F5 W
RS40*
Soft Carrying Case (Optional)
mA Input Fuse. F 500 mA, 250 V, (Fast blow), 1500 A breaking
capacity
10 A Input Fuse. F 15 A, 250 V, (Fast blow), 10,000 A minimum
breaking capacity; Or, F 15 A, 600 V, (Fast blow), 10,000 A minimum
breaking capacity 820829
Line Fuse. T 125 mA, 250 V, (Slow blow)
See Supplement
RS-232 terminal interface cable. Connects Fluke 45 to any properly
configured DTE connector (DB-25, male pins). IBM PC or IBM
PC/XT, IBM PS/2 (Models 25,30,50, 60,70,80).
838151
830802
822254