IEEE-488.2 Device Documentation Requirements C
C-3
MEAS?, MEAS1?, MEAS2?, VAL?, VAL1?, and VAL2? return one of two possible
formats (set with the FORMAT command). In addition, the VAL?, and MEAS?
queries will return two comma-separated values if both displays are in use:
Format 1: <NR3 NUMERIC RESPONSE DATA>
Format 2: <NR3 NUMERIC RESPONSE DATA> < UNIT>
MEASUREMENT <UNITS>
Volts dc VDC
Volts ac VAC
Current dc ADC
Current ac AAC
Resistance OHMS
Frequency HZ
Continuity Test VDC
Diode Test VDC
10. A description of any device-to-device message transfer traffic which does not follow
the rules for <RESPONSE MESSAGE> elements, Section 8.1.
There are no device-to-device messages.
11. The size of any block data responses, Section 8.7.9.4.
There are no block data responses.
12. A list of common commands and queries which are implemented, Section 10.
See Table 5-9 in the Users Manual.
13. A description of the state of the device after successful completion of the Calibration
query, Section 10.2.
The *CAL? command not implemented (an optional command).
14. The maximum length of the block used to define the trigger macro, if *DDT is
implemented, Section 10.4.
*DDT is not implemented.
15. The maximum length of macro labels, the maximum length of the block used to
define a macro, and how recursion is handled during macro expansion, if the macro
commands are implemented, Section 10.7.
Macros are not implemented.
16. A description of the response to the identification common query, *IDN?, Section
10.14.
The *IDN? query returns, for example:
FLUKE, 45, 1234567, 1.0 D1.0
where "1234567" will be replaced by the serial number of an actual meter, and "1.0
Dl .0 " are the version numbers of the main software and the display sub-system
software, respectively.
17. The size of the protected user data storage area, *PUD, Section 10.27.
*PUD not implemented. There is no protected user data storage area.