Fluke Ti40 Power Supply User Manual


 
Camera Setup
Setting Temperature Units
6
6-3
4. Position the pointer over the Display tab and tap E.
5. Position the pointer over the Temperature Scale box and tap E. A check
mark indicates function is enabled.
6. Tap the TRIGGER button to accept setting change and return to scan
target mode.
Setting Temperature Units
To set or change temperature units to Celsius, Fahrenheit, or Kelvin:
1. Use the MOUSE controller to position the pointer over the temperature
unit mark (ºF, ºK, ºC) on the task bar and tap E.
2. Repeat until the desired unit is selected.
Setting Temperature Calibration Range
You can change the temperature calibration range as needed: the narrower the
range the better the image quality, which enables you to see smaller
temperature variances. The wider the range the lower the image quality;
however, you are able to see a greater range of temperature variances.
To set the temperature calibration range:
Note
Calibration range cannot be changed on a paused or saved image. Be
sure to select the correct calibration range prior to pausing or saving
an image. See Appendix D to determine the calibrated temperature
ranges available for your camera model.
1. Tap F.
2. Use the MOUSE controller to position the pointer over Camera Settings
on the popup menu and tap E.
3. Position the pointer over the Lens tab and tap E.
4. Position the pointer over the calibration range radio button you want and
tap E.
5. Tap the TRIGGER button to accept setting change and return to scan
target mode.