Measurement Specialties 9116 Scanner User Manual


 
Pressure Systems, Inc. Model 9116 User’s Manual
Page 65 www.PressureSystems.com
READ TEMPERATURE COUNTS (Command ‘m’)
Purpose: Returns the most recently acquired raw temperature data for the specified
channels in averaged A/D counts (in the range -32768 to +32767). This
command is similar to command ‘a,’ except that the raw data reflects a channel’s
temperature signal instead of its pressure signal. Each datum returned in the
response will be in the specified high-precision data format, but representing A/D
counts as a signed integer average. This command is intended for advanced
users only and is not required for normal operation.
Command
mppppf
‘m’ is the command letter
pppp’ is the position field
f’ is the format field
Response
dddd.. dddd
dddd’ are the datum fields, each with a leading space (except f= 7
or 8).
Description: The 4-character hex position field (pppp) specifies a 16-bit binary bit-map, with
each bit (set to 1) specifying a particular channel number (16-1, left-to-right).
The 1-character format field (f) specifies the format of each data field (dddd) that
will be returned in the requested response. The first datum returned in the
response will be for the highest channel number supplied, and each (non-binary)
datum will be preceded by a space character. Some formats may not be
applicable to the specific type of data being requested. Valid formats are shown
in the following table:
f converts each internal response datum value from . . . max. char.
0 single binary float to 7-10-digit signed decimal “ [-xxx]x.xxxxxx” 13
1 single binary float to 8-digit hex integer “ xxxxxxxx” 9
2 double binary float to 16-digit hex integer “ xxxxxxxxxxxxxxxx” 17
5 single binary float to long integer (EU*1000) then to 8-digit hex integer 9
7 single binary float to single binary float (big endian: msb first) 4
8 single binary float to single binary float (little endian: lsb first) 4