Chapter 1 Installing and Configuring the SCXI-1127/1128
© National Instruments Corporation 1-19 SCXI-1127/1128 User Manual
• At least one other SCXI module other than the SCXI-1127/1128,
installed in the chassis
Consult the documentation for your SCXI chassis and other SCXI modules
for additional instructions and warnings. Any non-SCXI-1127/1128
modules should already be installed according to their installation
instructions. To set up the SCXI-1127/1128 module to be accessed by a
DAQ device via another SCXI module, follow these steps while referring
to Figure 1-9:
Figure 1-9.
Installing the SCXI-1127/1128 with Two DAQ Devices
You should have already installed the SCXI-1127/1128 modules and their
HVAB-backplane adapters (1357/1358) according to the instructions found
earlier in this chapter.
1. Connect the MIO/DIO cable adapter to the rear of a module that is not
the SCXI-1127/1128 module. Secure the adapter to the SCXI chassis
using the screws provided.
2. Attach your DAQ cable to the cable adapter at the rear connector.
1 Non-SCXI-1127/1128 Module
2 MIO or DIO Device
3 AB0+
4 AB0–
5 NI 4060 for PCI
6AB2+
7AB2–
8 SH9MD-9MD Cable
9 HV8-BAN4 Cable
10 2-Slot HVAB-Backplane Adapter
11 SCXI Chassis
12 Digital Signals Connector
13 SCXI-1127/1128 Module
14 HVAB Connector
1
14
11
12
13
2
9
8
10
7
6
5
4
3