NI PXI/PCI-5114 Specifications 16 ni.com
Environment
NI PXI-5114
Note To ensure that the NI PXI-5114 cools effectively, follow the guidelines in
the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit.
The NI PXI-5114 is intended for indoor use only.
Specification Value Comments
Operating
Temperature
0 ºC to +55 ºC in all NI PXI chassis except the following:
0 ºC to +45 ºC when installed in an NI PXI-1000/B or
PXI-101x chassis.
Meets IEC-60068-2-1 and IEC-60068-2-2.
—
Storage
Temperature
– 40 ºC to +71 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.
—
Operating Relative
Humidity
10% to 90%, noncondensing. Meets IEC-60068-2-56. —
Storage Relative
Humidity
5% to 95%, noncondensing. Meets IEC-60068-2-56. —
Operating Shock 30 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
—
Storage Shock 50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
—
Operating Vibration 5 Hz to 500 Hz, 0.31 g
rms
. Meets IEC-60068-2-64. —
Storage Vibration 5 Hz to 500 Hz, 2.46 g
rms
. Meets IEC-60068-2-64.
Test profile exceeds requirements of MIL-PRF-28800F,
Class B.
—
Altitude 2,000 m maximum (at 25 °C ambient temperature) —
Pollution Degree 2 —