© National Instruments Corporation 3 NI PXI/PCI-5114 Specifications
Voltage Levels
Full Scale (FS)
Input Range and
Programmable
Vertical Offset
50 Ω 1 MΩ
Range
(V
pk-pk
)
Vertical Offset
Range (V)
Range
(V
pk-pk
)
Vertical Offset
Range (V)
0.04 ±0.8 0.04 ±0.8
0.1 ±0.8 0.1 ±0.8
0.2 ±0.8 0.2 ±0.8
0.4 ±0.8 0.4 ±0.8
1 ±6.5 1.0 ±8.0
2 ±6.0 2.0 ±8.0
4 ±5.0 4.0 ±8.0
10 ±2.0
10 ±30
20 ±25
40 ±15
Maximum Input
Overload
50 Ω 1 MΩ
7 V
rms
with |Peaks| 10 V |Peaks| 35 V
Accuracy
Resolution 8 bits
DC Accuracy
(Programmable
Vertical
Offset = 0 V)
NI PXI-5114:
±(1.5% of Input + 0.3% of FS + 200 μV)
NI PCI-5114:
±(1.5% of Input + 0.3% of FS + 280 μV)
Within ±5 °C of
self-calibration
temperature.
Programmable
Vertical Offset
Accuracy
± 2% of offset setting Within ±5 °C of
self-calibration
temperature.
DC Drift ±(0.03% of Input + 0.06% of FS + 40 μV) per °C
Crosstalk,
Typical
–60 dB at 10 MHz
–45 dB at 100 MHz
CH 0 to/from
CH 1, External
Trigger to CH 0
or CH 1.
Specification Value Comments