NI PXI/PCI-5114 Specifications 6 ni.com
Horizontal
Sample Clock
Specification Value Comments
Sources Internal, Onboard Clock (internal VCXO)*
External, CLK IN (front panel SMB connector)
*
Internal Sample
Clock is locked to the
Reference Clock or
derived from the
onboard VCXO.
Onboard Clock (Internal VCXO)
Sample Rate
Range
Real-Time Sampling
(Single Shot)
Random Interleaved
Sampling (RIS)
*
Divide by n
decimation used
for all rates less than
250 MS/s. For more
information about
Sample Clock and
decimation, refer to
the NI High-Speed
Digitizers Help.
3.815 kS/s to 250MS/s* 250 MS/s to 5 GS/s in
increments of 250 MS/s
Timebase
Frequency
250 MHz When not using
External Sample
Clock.
Timebase
Accuracy
Not Phase-Locked to
Reference Clock
Phase-Locked to
Reference Clock
—
±25 ppm Equal to the Reference
Clock accuracy
Sample Clock
Delay Range
±1 Sample Clock period —
Sample Clock
Delay Resolution
≤ 20 ps —