National Instruments NI PXI/PCI-5114 Computer Hardware User Manual


 
© National Instruments Corporation 17 NI PXI/PCI-5114 Specifications
NI PCI-5114
Note To ensure that the NI PCI-5114 cools effectively, follow the guidelines in the
Maintain Forced Air Cooling Note to Users included in the NI PCI-5114 kit. Also, to
maximize airflow and extend the life of the device, leave any adjacent PCI slots empty.
The NI PCI-5114 is intended for indoor use only.
Specification Value Comments
Operating
Temperature
0 ºC to +45 ºC. Meets IEC-60068-2-1 and IEC-60068-2-2.
Storage
Temperature
– 40 ºC to +71 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.
Operating Relative
Humidity
10% to 90%, noncondensing. Meets IEC-60068-2-56.
Storage Relative
Humidity
5% to 95%, noncondensing. Meets IEC-60068-2-56.
Storage Shock 50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
Storage Vibration 5 Hz to 500 Hz, 2.46 g
rms
. Meets IEC-60068-2-64.
Test profile exceeds requirements of MIL-PRF-28800F,
Class B.
Altitude 2,000 m maximum (at 25 °C ambient temperature).
Pollution Degree 2