Appendix A Specifications
© National Instruments Corporation A-5 NI PXI-1052 User Manual
Operating Environment
Indoor use only
Ambient temperature range.................... 0 to 55 °C (Tested in accordance
with IEC-60068-2-1 and
IEC-60068-2-2.)
Operating relative humidity ................... 10 to 90%, noncondensing
(Tested in accordance with
IEC-60068-2-56.)
Altitude...................................................2,000 m (at 25 °C ambient
temperature)
Pollution Degree ....................................2
Storage Environment
Ambient temperature range....................–20 to 70 °C (Tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2.)
Storage relative humidity.......................5 to 95%, noncondensing
(Tested in accordance with
IEC-60068-2-56.)
Shock and Vibration
Operational shock ..................................30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)
Random Vibration
Operating ........................................5 to 500 Hz, 0.3 g
rms
Nonoperating ..................................5 to 500 Hz, 2.4 g
rms
(Tested in accordance with
IEC-60068-2-64. Nonoperating
test profile exceeds the
requirements of
MIL-PRF-28800F, Class 3.)