Philips V24CT Computer Monitor User Manual


 
Maintenance Checks
Monitor Installation and Patient Safety 11-15
Monitor Installation
and Patient Safety
This recommendation does not supersede any local procedures or legal
requirements.
Electrical Safety Tests At least every two years, after
system installation, after any
repairs, or as needed.
See “Testing and Maintaining the
CMS” in the “CMS Service Guide”
or “Testing and Maintaining the
V24 and V26“in the “V24 and V26
Service Guide”.
Testing of all cables and cords. Daily. See “Testing and Maintaining the
CMS” in the “CMS Service Guide”
or “Testing and Maintaining the
V24 and V26“in the “V24 and V26
Service Guide”.
Calibration of the measurement
modules where appropriate
(For EU/EFTA countries only)
Every 2 years. See individual module chapters in
Volume 2 of this manual for
details.
User Tests Module dependent. Please refer to
chapter 13 “Maintenance” of this
manual or chapter 3 of the CMS
Service Guide “Maintaining the
CMS” for details.
See chapter 13 “Maintenance” of
this manual for details.
Maintenance Frequency Source of Information