Soehnle 3020 Postal Equipment User Manual


 
English
88
2. Basic functions
The basic functions are similar or identical for all application programs.
2.1 Power on/off
Power-on
After power-on and during the run-up phase, the Soehnle Professional logo is
displayed.
The measuring points are briefly listed as they are detected.
The scale is zeroed after the power-on routine. (This may vary with
non-approved calibration). On power-on, the last application program
activated is displayed.
Zero range limits at power-on
Approvable: Zero range at power-on 20% of the weighing range.
The default value is -5% to +15% of the weighing range.
Non-approved: Zero range at power-on -99% to +99% of the weighing range.
If the appliance is switched on outside the zero range, the scale displays error
message Error 50 / 51 Value outside zero range limit.
When the cause for overload or underload is rectified, the scale is zeroed without
having to switch the appliance on and off and is ready to operate.
Response to power failure
The scale returns to the previous settings when power is restored.
Power-off
Press the On/Off key.
If the activated measuring point is loaded or a menu is open, the scale switches
off only 3 seconds after the On/Off key is pressed.
2.2 Selecting a program
Select one of the application programs from the list.
Call up the setting menu by pressing the Print key and F key
(press simultaneously).
• The arrow points to "Program Settings" .
• Press the Print key: the arrow points to "General".
• Press the Print key: the arrow points to "Programs".
• Press the Print key: the arrow points to the program last selected.
• Select the program you wish to activate by pressing the Tare key or F key.
• Confirm the program by pressing the Print key.