Chapter 2 Performance Tests
58502A Operational Verification
User’s Guide 2-7
Alarm and Status Output Verification
1 Connect the equipment as shown in Figure 2-2, using output 12 as the
test output. (Note that any one of the outputs can be used for this
alarm verification.)
Figure 2-2. Alarm/Status Output Verification Setup
2 Verify the ALARM/STATUS output function by shorting to ground any
of the outputs (1-12). Let’s use output 12 as shown in Figure 2-3.
Figure 2-3. Output 12 Grounded
3 Observe that the front-panel Alarm indicator illuminates and the
ALARM/STATUS output pins (2, 3, and 4) go active (TTL low,
approximately 20 mV).
Figure 2-4 shows the output pins (2, 3, and 4) and ground pins (1, 5,
and 9.