Xantrex Technology XDC Power Supply User Manual


 
Operation
Power Supply Operation
Release 3.0 69
To set the Under-Current Protection:
To set the Over-Power Protection:
Step # Do This You Will See
1 OVP SET 0V
2 Press repeatedly until the
desired setting appears.
UCP SET 0A
3 UCP SET 0A
4
Use the Current knob, the numeric
keypad, or the arrow keys to enter a
value. The value must be between
0–103% of the unit’s rated current.
UCP SET ####A
5 S/D if trip? N
6
Use the arrow keys to select Yes or No.
For this example, Yes is selected.
S/D if trip? Y
7 This setting is saved and the display
returns to its default operating mode.
Step # Do This You Will See
1 OVP SET 0V
2 Press repeatedly until the
desired setting appears.
OPP SET 0W
3 OPP SET 0W
4
Use both the Current and Voltage
knobs, or the numeric keypad, or the
arrow keys to enter a value. The
value must be between 0W and
103% of the unit’s rated power.
OPP SET ####W
5 S/D if trip? N
6 Use the arrow keys to select Yes or
No. For this example, Yes is selected.
S/D if trip? Y
7 This setting is saved and the display
returns to its default operating mode.
PROT SET
ALARMS
PROT SET
ALARMS
ENTER
ENTER
ENTER
PROT SET
ALARMS
PROT SET
ALARMS
ENTER
ENTER
ENTER