A SERVICE OF

logo

C-8 DATA COMPUTATIONS
MON2000
Conc. Analysis with Response Factor JULY 2010
Component concentrations may be input
through analog inputs 1 to 4 or may be fixed. If
a fixed value is used, the calibration for that
component is the mole percent that will be used
for all analyses:.
where
Equation C-5 Concentration Calculations with Normalization
CONCN
n
normalized concentration of component
“n” in percent of total gas concentration
CONC
n
non-normalized concentration of
component “n” in mole percent
CONC
i
non-normalized concentration (in mole
percent) from each of the “k” components
to be grouped into this normalization
k number of components to be included in
the normalization
CONCN
n
CONC
n
CONC
i
i 1=
k
----------------------------
100×=