A&D AD-4212A-100/200/600/1000 Postal Equipment User Manual


 
29
10-3 Details of the Function Table
Class Item and Parameter Description
0
1
Cond
Condition
2
Fast weighing speed, sensitive value
FAST
MID.
Slow weighing speed, stable value
SLOW
0
1
5t-b
Stability band width
2
Stable range is ±1 digit
Stable range is ±3 digits
The stabilization indicator
illuminates when the display
fluctuation is within the range.
0
OFF
Hold *1
Hold function
1
ON
Holds the display when stable in
animal mode. With “Hold 1”,
ANIMAL turns on.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by tracking
zero drift.
0
5 times/second
5pd
Display refresh rate
1
10 times/second
Period to refresh the display
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal point format
0
OFF
p-on
Auto display-ON
1
ON
Turns on the weighing mode
display when the AC adapter is
plugged in.
0
OFF
poff
*1
Auto display-OFF
1
ON (10 minutes)
Turns off the display after 10
minutes of inactivity.
0
OFF
g5i
Capacity indicator
1
ON
Capacity indicator
Zero: 0%,
Maximum capacity: 100%
0
Displays the minimum weighing
value
1
Does not display the minimum
weighing value
Factory setting
1: AD-4212B-101/102
0: Other models
rng
Minimum weighing
value display at start
2
Retains the previous condition
0
Does not tare (Displays the previous value)
ba5fnc
Environment
Display
p-tr
Tare at start
1
Tares (Displays zero)
Cl adj
Clock
Refer to "10-9 Clock and Calendar Function".
The time and date are added to
the output data.
Factory setting
Note: “Digit” is a unit of minimum weighing value
*1 Only the AD-4212A series balance displays this item.