A&D AD-4212A-100/200/600/1000 Postal Equipment User Manual


 
78
When the balance is built into a system, remove the balance from the system. Place it on a
solid table. Install the breeze break and perform checking.
When the balance proper performance is confirmed, refer to page 10 to set up the installation
site.
Checking that the operating environment or weighing method is proper
Operating environment
Is the weighing table solid enough
Is the balance level? Refer to “3-1 Before Use”.
Is the operating environment free from vibration and drafts? Has the stainless steel breeze
break been installed?
Is there a strong electrical or magnetic noise source such as a motor near the balance?
Is there a heat source near the balance?
Weighing method
Does the weighing pan touch the breeze break or anything? Is the weighing pan installed
correctly?
Is the RE-ZERO key pressed before placing a sample on the weighing pan? Or, is RE-ZERO
performed using external contact input or an RS-232C command?
Is the sample placed in the center of the weighing pan?
Has the balance been warmed up for the appropriate duration before weighing?
AD-4212A series balance: 30 minutes or more
AD-4212B series balance: one hour or more
Are the leveling feet of the weighing unit placed flat to the installation surface?
If not, the weight value will be unstable or the specified repeatability can not be obtained.
Improve the installation condition, by securing the weighing unit or reducing the vibration that is
conveyed to the weighing unit.
Sample and container
Has the sample absorbed or lost moisture due to the ambient conditions such as temperature
or humidity?
Has the temperature of the container been allowed to equalize to the ambient temperature?
Refer to “3-2 When Building into a System”.
Is the sample charged with static electricity? Refer to “3-2 When Building into a System”.
Is the sample of magnetic material such as iron? Use much care when weighing magnetic
materials. Refer to “3-2 When Building into a System”.