10-4 Description of the Class “Environment, Display”
Condition ( Cond )
Cond 0 This parameter is for sensitive response to the fluctuation of a weight value.
Used when fast weighing speed is required.
After setting, the balance displays FAST.
Cond 2 This parameter is for stable weighing with slow response. Used to prevent a
weight value drift due to vibration or drafts.
After setting, the balance displays SLOW.
Note
For the AD-4212A series balance, with “Hold function (Hold)” set to “ON
(1)”, this item is used to set the averaging time.
Stability band width ( 5t-b )
This item controls the width to regard a weight value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs or
stores the data. The parameter influences the “Auto print mode”
5t-b 0 This parameter is for sensitive response of the stabilization indicator. Used
for exact weighing.
5t-b 2 This parameter ignores slight fluctuations of a weight value. Used to prevent
a weight value from drifting due to vibration or drafts.
Note
For the AD-4212A series balance, with “Hold function (Hold)” set to
“ON (1)”, this item is used to set the stabilization range..
Hold function ( Hold ) (Animal weighing mode, AD-4212A only)
This function is used to weigh a moving object such as an animal.
When the weight value is over the weighing range from zero and the display fluctuation is within the
stabilization range for a fixed period of averaging time, the processing indicator illuminates and the
balance displays the average weight of the animal. When the animal is removed from the weighing
pan, the display returns to zero automatically.
This function is available only when the hold function parameter is set to “1 (the animal mode
indicator ANIMAL illuminates) and any weighing unit other than the counting mode is selected.
The averaging time and stabilization range are set in “Condition (Cond)” and “Stability band width (5t-b)”.
Weighing range Averaging time Stabilization range
Cond 0 5t-b 0
Cond 1 5t-b 1
0.2 g or over
Cond 2
2 seconds Faster
4 seconds |
8 seconds More accurate
5t-b 2
Small
|
Big