31
Class Item and Parameter Description
0
Plus only
Displayed value>Reference
1
Minus only
Displayed value<Reference
ap-p
Auto print polarity
2
Both
Regardless of displayed value
0
10 digits
1
100 digits
ap-b
Auto print difference
2
1000 digits
Difference between reference
value and displayed value
0 Not used
1
Stores unit mass in counting mode
2
Stores weighing data
3
Stores calibration data
data
Data memory
4
Stores upper/lower limit values
Related items:
prt, int, d-no, 5-td, info
0
Every measurement
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minute
6
2 minutes
7
5 minutes
int
Interval time
8
10 minutes
Interval time in the interval
memory mode when using
prt 3,
data 2
0
No output
d-no
Data number output
1
Output
Refer to "14. DATA MEMORY".
0
No output
1
Time only
2
Date only
5-td
Time/Date output
3
Time and date
Selects whether or not the time
or date is added to the weighting
data. For details, refer to "10-9
Clock and Calendar Function".
0
No output
5-id
ID number output
1
Output
Selects whether or not the ID
number is output.
0
No pause
pU5e
Data output pause
1
Pause (1.6 seconds)
Selects the data output interval.
0
Not used
at-f
Auto feed
1
Used
Selects whether or not automatic
feed is performed.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
For how to set time and date to
be added, refer to “10-9 Clock
and Calendar Function”.
0
Not used
dout
Data output
Not displayed
when BCD
output (OP-01)
is installed.
ar-d
Zero after output
1
Used
Adjusts zero automatically after
data is output.
Factory setting
Note: “Digit” is a unit of minimum weighing value