Calibration kits
Error correction requires that the systematic
errors in the measurement system be characterized
by measuring known devices (standards) over the
frequency range of interest with the process of cal-
ibration. All calibration kits contain standards
used for this purpose. The standards in the 3.5 mm,
2.4 mm, and Type-N calibration kits use the preci-
sion slotless connector (PSC-3.5, PSC-2.4, and
PSC-N). Unless otherwise noted all coaxial calibra-
tion kits include connector gauges and a torque
wrench. Option 002 provides calibration kit data
on magnetic tape for use with the 8510A/B (not
available with 85059A).
Mechanical Connector Frequency Description
calibration Kit type range (GHz)
85050B 7 mm 0.045 to 18 Contains open and short circuits, fixed and sliding terminations.
85050C 7 mm 0.045 to 18 Precision kit. Contains standards for TRL calibration, including precision airline. Also
contains open and short circuits and fixed termination.
85050D 7 mm 0.045 to 18 Economy kit. Contains open and short circuits and precision fixed termination. Gauges
not included.
85052B 3.5 mm 0.045 to 26.5 Contains open and short circuits, fixed and sliding terminations, and in-series adapters.
85052C 3.5 mm 0.045 to 26.5 Precision kit. Contains standards for TRL calibration, including precision airlines. Also
contains open and short circuits, fixed terminations, and in-series adapters. Gauges not
included.
85052D 3.5 mm 0.045 to 26.5 Economy kit. Contains open and short circuits, precision fixed termination, and in-series
adapters. Gauges not included.
85054B Type-N 0.045 to 18 Contains open and short circuits, fixed and sliding terminations, in-series adapters, and
7 mm-to-Type-N adapters.
85054D Type-N 0.045 to 18 Economy kit. Contains open and short circuits, fixed terminations, in-series adapters and
7 mm-to-Type-N adapters. Gauges not included.
85056A 2.4 mm 0.045 to 50 Contains open and short circuits, fixed and sliding terminations, and in-series adapters.
85056D 2.4 mm 0.045 to 50 Economy kit. Contains open and short circuits, fixed terminations, and in-series adapters.
Gauges not included.
85056K 2.92/2.4 mm 0.045 to 50 Contains 2.4 mm open and short circuits, fixed loads, and 2.92 mm adapters.
85059A 1.0 mm 0.045 to 110 Broadband coaxial precision calibration kit consists of 1.0 mm shorts, opens, fixed loads, and
in-series adapters. It also includes offset-shorts covering 50 to 110 GHz. Gauges not included.
11904S 2.92 mm 0.045 to 40 Must be used with 85056A/D 2.4 mm calibration kit. Includes four 2.92-mm-to-2.4-mm
adapters. Gauges not included.
X11644A WR-90 8.2 to 12.4
P11644A WR-62 12.4 to 18
K11644A WR-42 18.0 to 26.5
R11644A WR-28 26.5 to 40
Q11644A WR-22 33 to 50
U11644A WR-19 40 to 60
V11644A WR-15 50 to 75
W11644A WR-10 75 to 110
Contains standards for TRL calibration. Includes precision wave-
guide section, short circuit, and fixed or sliding terminations.