Agilent Technologies E6701A Network Router User Manual


 
Copyright 2002 Agilent Technologies. 10001810-E6701A Tutorial LAN Connection
All Rights Reserved.
Page 29 of 32 Rev. 1.2; Feb. 14, 2002
Parameters table to LLC Frame Check Sequence; change the table value to
Corrupt.
2.2.3. It is possible that the mobile is capable but GMM is prioritised too low and
the wireless appliance is unable to sustain the link.
2.2.3.1. To get around this we can change the BLER block polling period. This
will allow the mobile to send the ACK/NACK less frequently(one negative
result: TX measurements will be slower since the mobile is not transmitting
every frame.) Go to call parameters, 1 of 4.
3. ETSI Test Mode measurements have problems:
3.1. Check that the mobile is capable of the ETSI test mode you would like to use, if not it
will not work
3.2. Check the same parameters as in Attach problems.
4.
If the attach condition is terminated, this may be a Connection
Reconfiguration problem:
4.1. This may be a Connection Reconfiguration problem. When any of the connection
parameters are reconfigured the connection is terminated.
4.2. The mobile likely does not have the PACKET_TIMESLOT_RECONFIGURE
message implemented.
4.2.1. This message is used to change MSTXL, channel, Band, Coding Scheme,
Multislot configuration, and offset P0.
4.2.2. The solution when using the PACKET_TIMESLOT_RECONFIGURE
message is to stop the data connection before changing parameters, then restart the
data connection.
4.3. Alternatively, you can change the message used to change parameters to
PACKET_DOWNLINK_ASSIGNMENT; or, for power changes use the
PACKET_POWER_AND_TIMING_ADVANCE message. Go to call parameters, 1 of
4.
5. If Internet Explorer cannot access the 8960 Welcome Page:
5.1. Connections:
5.1.1. Check that the LAN cables are good, and are properly connected
5.1.2. If you are using the Test Sets front panel DATA port: check that the LAN
jumper cable is attached on the Test Set rear panel between the LAN PORT
connector and the ETHERNET TO FRONT PANEL connector. (NOTE: For
minimum RF interference, Agilent recommends using the rear-panel LAN PORT
connector).
5.2. Test Set:
5.2.1. On the System Config Screen, check that the Application is GPRS Lab
Application.
5.2.2. The Test Set LAN IP Address and Subnet Mask should not include leading
zeros.
Example: 111.22.333.44 is correct
111.022.333.044 is incorrect