Agilent Technologies HP 16505A Network Card User Manual


 
Procedure
This procedure verifies the performance of the interface module. To verify
performance of the HP 16500B Logic Analysis System or the optional
measurement modules, refer to the Service Guides for those products.
1
If the screen appears blank, the interface module or cable may be
defective.
a
Remove power from the HP 16500B, disconnect the power cable, then
remove the top cover.
b
Remove and inspect the interface module cable for signs of damage.
If the cable appears defective, replace it.
c
Ensure that the cable is connected according to the installation
procedures and that all cables are properly seated. For installation
procedures, refer to chapter 2, "Installing the Interface Module."
d
Install the top cover of the HP 16500B mainframe, connect the power
cable, and reapply power.
2
Touch the blue field in the upper-left corner of the HP 16500B screen.
Select
System
from the pop-up menu .
3
Touch the blue field immediately to the right. In the pop-up menu,
select
Configuration
.
4
Verify that "Ethernet" is available in the Communications menu on
the HP 16500B System Configuration screen.
Ethernet should be
listed here
Servicing the Interface Module
Verify the Interface Module Performance
3-3