Pressure Systems 9816 Scanner User Manual


 
Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©
Page 43 www.PressureSystems.com
RESET (Command 'B')
Purpose: Instructs the module to reset internal operating parameters, and to set all internal
control variables to their default “reset” state (see description below). The current
TCP/IP socket connection will remain open. Execution after a power off/on cycle
is optional (unnecessary).
Command
B
‘B’ is the command letter
Response
A
‘A’ is the acknowledge letter
Description: The module returns to the following “reset” states if this command is executed:
Re-zero correction (offset) terms are set to the last values stored in
transducer memory.
Span correction (gain) terms are set to the last values stored in transducer
memory.
Calibration Valve is set to the RUN Position.
Number of Samples for Data Averaging is set to last value stored in non-
volatile memory (factory default = 8).
Any autonomous host data delivery streams defined by ‘c’ sub-commands
are reset (undefined).
The Multi-Point Calibration function defined by ‘C’ sub-commands is reset
(undefined) if in progress.
Example:
Send TCP/IP command to a Model 9816 module (via open socket) to reset defaults:
B
Read following response:
"A"