Fluke A20M-AF Computer Monitor User Manual


 
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company Page 2
1.2 SPECIFICATIONS
1.2.1 GENERAL SPECIFICATIONS
Power Requirements: 85 to 264 VAC, 50/60 Hz, 25 VA max consumption
and 12VDC, 1.2 A
Operating Temperature Range: 18 to 28 °C
Storage Temperature Range: - 20 to 70 °C
Weight: 11 kg (22 lb) approx.
Dimensions: 23.5 cm H x 29.4 cm W x 37.5 cm D (9.3 in. x 11.6 in. x 14.8 in.)
Communication Ports
Fuses:
RS232 (COM1, COM2), IEEE-488.2
1 A, 250 VAC fuse, 5 x 20 mm, time lag type fuse
Internal power supply fuse not replaceable by operator: 2.5A, 250 VAC
Pressure Range: 0 to 10 000 psi (70 MPa) in gauge mode
Atmosphere to 10 000 psi (70 MPa) in absolute mode
Operating Medium: Any clean, dry, non-corrosive gas
Pressure Connections: DH500 female
DH500 is a gland and collar type fitting for 6mm (1/4 in.) coned and left
hand threaded tubes equivalent to AE F250C, HIP HF4, etc.
Pressure Limits: Maximum Working Pressure: 10 000 psi (70 MPa)
Maximum Pressure Without Damage: 12 000 psi (83 MPa)
When low Q-RPT is connected to pressure:
Maximum Working Pressure: 3 000 psi (20 MPa)
Maximum Pressure Without Damage: 3 600 psi (24 MPa)
PGC-10000-AF SYSTEM
Temperature: Operating:
18 to 28 °C
Storage:
-20 to 70 °C
Relative Humidity: Operating: 15 to 70%RH (non-condensing)