Fluke A20M-AF Computer Monitor User Manual


 
3. OPERATION
Page 37 © 2007 DH Instruments, a Fluke Company
3.3.6.6 Freeze
PURPOSE
To activate the Freeze DISPLAY.
See Section 3.3.6, Principle.
OPERATION
To activate the Freeze DISPLAY press [DISPLAY], <6Freeze>. Pressing
<6Freeze> returns to the main run screen with the Freeze DISPLAY active.
With the Freeze DISPLAY active the main run screen is:
1. Pressure measured by active range of RPM4
when [ENT] was pressed (displays 0.00 by
default when Freeze DISPLAY is first
activated).
4435.6 psi a zHi
F 4420.7
Pressing [ENT] causes the current pressure measured by the active RPM4 Q-RPT
to be captured and displayed.
If the pressure measurement unit, mode or range is changed while the Freeze
DISPLAY is active, the Freeze value defaults back to zero.
To go to a DISPLAY other than Freeze, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).
3.3.6.7 Clean
PURPOSE
To activate the Clean DISPLAY.
See Section 3.3.6, Principle..
OPERATION
To activate the Clean DISPLAY press [DISPLAY], <7Clean>. Pressing <7Clean>
returns to the main run screen with the Clean DISPLAY active.
With the Clean DISPLAY active the main run screen is:
1. Conventional main run screen first line.
2. “Clean” second line.
99.76 psi a zHi
To go to a DISPLAY other than Clean, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).