National Instruments 1192 Switch User Manual


 
Chapter 1 Installing and Configuring the SCXI-119X
SCXI-1190/1191/1192 User Manual 1-12 ni.com
e. Change the Connected to selection from None to Connected to:
Device x.
If you get the warning message
Failed to find
followedbythe
module codes and the message
Unable to communicate with
chassis
, take the following troubleshooting actions:
a. Make sure that the SCXI chassis is powered on.
b. Make sure the cable between the SCXI chassis and DAQ device or
computer-based instrument is properly connected.
c. Inspect the cable and the connectors for any bent pins.
d. Make sure you are using the correct National Instruments cable
assembly.
e. Test your DAQ device or computer-based instrument to verify it is
working properly. Refer to your DAQ device or computer-based
instrument user manual for more information.
If you get the warning message
Failed to find
, followed by
module codesand thenthe message
Instead found: module with
ID 0Xxx
:
a. Return to the Connecting the SCXI-119X to a DAQ Device or
Computer-Based Instrument section and make sure the correct
module is in the specified slot.
b. Delete the incorrect module as described in the Removing the
SCXI-119X from an SCXI Chassis section.
c. Add the correct module as described in the Manually Adding
Modules section.
If you get the warning message
Failed to find
,followedbya
module code and then the message
Slot x is empty
:
a. Check to see if the configured module is installed in the specified
slot. If not, install the module by referring to the Installing the
Hardware section.
b. If the module is installed in the correct slot, turn off the chassis,
and remove the module as specified in the Removing the
SCXI-119X from an SCXI Chassis section. Verify that no
connector pins are bent on the rear signal connector. Reinstall the
module, ensuringthe module isfully inserted andproperly aligned
in the slot.
After checking the preceding items, return to the Self-Test Verification
section and retest your SCXI chassis.