National Instruments VXI-USB Network Card User Manual


 
Appendix A Specifications
VXI-USB User Manual A-2 ni.com
Shock and Vibration
Operational shock...................................30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)
Random vibration
Operating.........................................5 to 500 Hz, 0.3 g
rms
Nonoperating...................................5 to 500 Hz, 2.4 g
rms
(Tested in accordance with
IEC-60068-2-64. Nonoperating
test profile exceeds the
requirements of
MIL-PRF-28800F, Class 3.)
Power Requirement
+5 V (excluding downstream USB devices)
Typical.............................................2 A
Maximum ........................................4 A
–5.2 V
Typical.............................................200 mA
Maximum ........................................500 mA
–2 V
Typical.............................................100 mA
Maximum ........................................250 mA
Physical
Size .........................................................C size, C-1
Dimensions ............................................. 23.3 × 43.0 cm (9.2 × 13.4 in.)
Weight ....................................................1.14 kg (2.5 lb)
I/O connectors
USB device (Series A).....................2
USB host (Series B).........................1
SMB.................................................3