Oki C9800 Printer User Manual


 
2
2-49 Using the EFI Densitometer ED-100
C9800 Color Guide - 49
9. Place the ED-100 over the specified patch on the Measurement Page, making sure to
center the sample aperture over the patch. Press Measure to take the measurement.
N
OTE: For a more accurate measurement, place
several plain sheets of paper beneath the
Measurement Page to block underlying colors
from being read by the instrument.
If you have specified multiple clicks per patch,
continue pressing Measure to take the required
number of measurements. We recommend that
you move the ED-100 slightly after each
measurement while keeping it within the patch
circle.
When the patch has been measured successfully, a
check mark appears in the patch preview in the
Measurements window. The window identifies
and highlights the next patch to be measured.
10. Repeat step 9 to measure each successive patch on the Measurement Page.
Start with A1 and continue down Column A. When you have finished measuring all
the patches in Column A, proceed to Column B in the same manner until all columns
have been measured.
To retake a measurement, select the desired patch preview in the Measurements
window and measure the patch on the page using the ED-100.
NOTE: You must take the measurements in the order outlined in the Measurements
window and on the measurement page.
11. When all the patches have been read successfully, click Accept.
12. Click Apply in the Calibrator window to implement the new calibration set.
13. Click OK in the Information dialog box.
N
OTE: This completes the Fiery calibration process.