Oki C9800 Printer User Manual


 
A
A-88 Color Measurement Instruments
C9800 Color Guide - 88
Multiple users can be connected to one Fiery with ColorWise Pro Tools, but only one
user at a time can use the Calibrator module. An error message appears if you try to
calibrate when another user is already using ColorWise Pro Tools to calibrate.
You will need the Color Reflection Reference strip that is included with the
spectrophotometer. Remove the strip from its protective envelope, handling it by the
edges only. Keep the strip free of dust and smudges and always store it in the protective
envelope.
TO CALIBRATE THE X-RITE DTP41
1. Connect the spectrophotometer to the computer and turn on power (see page A-84).
2. Start Calibrator.
For instructions on starting Calibrator, see page 2-38.
3. Choose DTP41 as the measurement method.
4. In the Get Measurements pane, click Measure.
5. In the Measurement Options dialog box, click Measure.